Z-2146-2014 Class II Terminated
Recalled by Brainlab AG
— Feldkirchen, N/A
FDA device recall Z-2146-2014 was initiated by Brainlab AG on Jun 24, 2014 and is designated Class II. Reason for recall: When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite… The recall status is terminated (terminated Apr 24, 2017). Affected quantity: 15 systems.
Recall details
- Product Type
- Devices
- Report Date
- Aug 13, 2014
- Initiation Date
- Jun 24, 2014
- Termination Date
- Apr 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 systems
Product description
ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment.
Reason for recall
When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.
Distribution pattern
Worldwide Distribution - USA including CA, CT, and IL; Internationally to France, Germany, Japan, Netherlands, Russia, and Slovenia.
Code information
ExacTrac Version 6.0.4; Model/Catalog Numbers: 1) 20833C EXACTRAC 6.0 IR POSITIONING SOFTWARE; 2) 20834C EXACTRAC 6.0 X-RAY POSITIONING SOFTWARE; 3) 20835C EXACTRAC 6.0 IR MONITORING SOFTWARE; 4) 20853C EXACTRAC 6.0 X-RAY LIMITED SOFTWARE; 5) 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR); 6) 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR); 7) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR); 8) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED; 9) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC; 10) 49936A ET SOFTWARE UPDATE 6.0.X TO 6.0.4; 11) 49938 ET SOFTWARE UPDATE 6.0.3TO 6.0.4; 12) 49939 ET SOFTWARE UPDATE 6.0.X to 6.0.4 P&R; 13) 49973B ET DATA PREP / REVIEW SYSTEM; 14) 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY; 15) 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED; 16) 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2146-2014
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Brainlab AG? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works