Z-2147-2012 Class II Terminated

Recalled by Linvatec Corp. dba ConMed Linvatec — Largo, FL

Recall Details

Product Type
Devices
Report Date
August 15, 2012
Initiation Date
August 12, 2011
Termination Date
November 21, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product Description

CrossFT Punch, Product Number PFT-45M. Manufacturer: CONMED LINVATEC. The intended use is to create a hole to facilitate insertion of the CrossFT suture anchor into bone.

Reason for Recall

ConMed Linvatec recalled lot number #156920, PFT-45M, CrossFT Punch because there is a possibility that the device was etched incorrectly listing the item number as PFT-00M and includes incorrect depth lines for insertion.

Distribution Pattern

Nationwide Distribution, including the states of Colorado and Pennsylvania.

Code Information

Lot #156920