Z-2173-2013 Class II Terminated

Recalled by Roche Molecular Systems, Inc. — Branchburg, NJ

FDA device recall Z-2173-2013 was initiated by Roche Molecular Systems, Inc. on Jul 25, 2013 and is designated Class II. Reason for recall: During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas¿ 4800 HPV Master Mix reagent that was packaged in four (4) cobas¿ 4800 AMP/DET… The recall status is terminated (terminated Aug 6, 2015). Affected quantity: US-IVD: 350 Kits; CE-IVD: 441 KITS.

Recall details

Product Type
Devices
Report Date
Sep 18, 2013
Initiation Date
Jul 25, 2013
Termination Date
Aug 6, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US-IVD: 350 Kits; CE-IVD: 441 KITS

Product description

KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 South Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 USA The cobas¿ HPV Test is a qualitative in vitro test for the detection of Human Papillomavirus in patient specimens. The test utilizes amplification of target DNA by the Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the detection of 14 high-risk (HR) HPV types in a single analysis. The test specifically identifies types HPV16 and HPV18 while concurrently detecting the rest of the high risk types (31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68). The cobas¿ HPV Test is indicated: (a) To screen patients 21 years and older with ASC-US (Atypical squamous cells of undetermined significance) cervical cytology test results to determine the need for referral to colposcopy. (b) To be used in patients 21 years and older with ASC-US cervical cytology results, to assess the presence or absence of high-risk HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. The results of this test are not intended to prevent women from proceeding to colposcopy. (c) In women 30 years and older, the cobas¿ HPV Test can be used with cervical cytology to adjunctively screen to assess the presence or absence of high risk HPV types. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. (d) In women 30 years and older, the cobas¿ HPV Test can be used to assess the presence or absence of HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. Cervical specimens that may be tested with the cobas¿ HPV Test include the following liquid based collection media and collection device: " ThinPrep¿ Pap TestTM PreservCyt¿ Solution " Endocervical Brush/Spatula

Reason for recall

During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas¿ 4800 HPV Master Mix reagent that was packaged in four (4) cobas¿ 4800 AMP/DET kits, US- and CE-IVD (240 and 960 test kit configurations). Examination of the batch production record of the material revealed there was an error in the tare weight used for the carboy during the fina

Distribution pattern

Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Mexico, Italy, New Zealand, Netherlands, Peru, Norway, Singapore, Romania, South Africa, Spain, South Korea, Sweden, Switzerland and UK.

Code information

US-IVD: Material number 05232880190 Lot Number R13814, R15785 CE-IVD: Material Number: 05235898190 05235898190 05235901190 05235910190 Lot Number R15771, R15304

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2173-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2173-2013. Report a data issue: hello@regdatalab.com.