Z-2198-2021 Class II Terminated

Recalled by Straumann USA LLC — Andover, MA

FDA device recall Z-2198-2021 was initiated by Straumann USA LLC on May 27, 2021 and is designated Class II. Reason for recall: Tip of the screwdriver does not have the correct configuration / profile therefore the screwdriver does not properly fit the opposing screw head and cannot pick up the screws, may result in injury The recall status is terminated (terminated Mar 21, 2023). Affected quantity: 130 pieces.

Recall details

Product Type
Devices
Report Date
Aug 11, 2021
Initiation Date
May 27, 2021
Termination Date
Mar 21, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
130 pieces

Product description

Straumann SCS Screwdriver, driver, for ratchet, short, L 21mm, stainless steel Article Number: 046.401. dental surgical instrument.

Reason for recall

Tip of the screwdriver does not have the correct configuration / profile therefore the screwdriver does not properly fit the opposing screw head and cannot pick up the screws, may result in injury

Distribution pattern

US Nationwide Distribution

Code information

Packaged lot number: EHG85 Lot Number engraved on part: CXV34

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2198-2021
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2198-2021. Report a data issue: hello@regdatalab.com.