Z-2209-2012 Class II Terminated

Recalled by Roche Molecular Systems, Inc. — Branchburg, NJ

FDA device recall Z-2209-2012 was initiated by Roche Molecular Systems, Inc. on Feb 1, 2012 and is designated Class II. Reason for recall: Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument… The recall status is terminated (terminated Jul 16, 2013). Affected quantity: 2,132 boxes.

Recall details

Product Type
Devices
Report Date
Aug 22, 2012
Initiation Date
Feb 1, 2012
Termination Date
Jul 16, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,132 boxes

Product description

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

Reason for recall

Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce

Distribution pattern

Nationwide Distribution (USA).

Code information

Lot Numbers: 11049503, 11053503, 11054503, 11525503 and 11526503.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2209-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2209-2012. Report a data issue: hello@regdatalab.com.