Z-2267-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- August 13, 2025
- Initiation Date
- June 25, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 system
Product Description
DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Reason for Recall
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Distribution Pattern
Worldwide distribution - US Nationwide.
Code Information
UDI-DI: 15099590369231; Serial Numbers: 608109