Z-2394-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 8, 2021
- Initiation Date
- August 10, 2021
- Termination Date
- May 17, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 cases (1/case)
Product Description
TRUMATCH CT PIN GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420578
Reason for Recall
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Distribution Pattern
Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.
Code Information
Lot Numbers: 27932 , 27806, 27515, 27466, GTIN: 10603295430797