Z-2398-2021 Class II Terminated

Recalled by DePuy Orthopaedics, Inc. — Warsaw, IN

FDA device recall Z-2398-2021 was initiated by DePuy Orthopaedics, Inc. on Aug 10, 2021 and is designated Class II. Reason for recall: Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic… The recall status is terminated (terminated May 17, 2023). Affected quantity: 5 cases ( 1/cs).

Recall details

Product Type
Devices
Report Date
Sep 8, 2021
Initiation Date
Aug 10, 2021
Termination Date
May 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 cases ( 1/cs)

Product description

TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420915

Reason for recall

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

Distribution pattern

Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.

Code information

Lot number/ GTIN: 28289, 28247, 27876, 27432, 27613 (OUS) GTIN: 10603295384854

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2398-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for DePuy Orthopaedics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for DePuy Orthopaedics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about DePuy Orthopaedics, Inc. →

Source: openFDA record for Z-2398-2021. Report a data issue: hello@regdatalab.com.