Z-2403-2012 Class II Terminated

Recalled by Roche Molecular Systems, Inc. — Branchburg, NJ

FDA device recall Z-2403-2012 was initiated by Roche Molecular Systems, Inc. on Apr 23, 2012 and is designated Class II. Reason for recall: When using a Laboratory Information system (LIS) with automated systems utilizing AMPLILINK software v3.2 series (AL v3.2) there is a potential for wrong results to be assigned to an order. The recall status is terminated (terminated Feb 12, 2014). Affected quantity: 1292 instruments.

Recall details

Product Type
Devices
Report Date
Sep 26, 2012
Initiation Date
Apr 23, 2012
Termination Date
Feb 12, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1292 instruments

Product description

Amplilink (AL) Software 3.2.3 for use with COBAS AMPLICOR, CAP/CA, COBAS TaqMan,and COBAS AmpliPrep/COBAS TaqMan. The software can be used with a Laboratory Information System.

Reason for recall

When using a Laboratory Information system (LIS) with automated systems utilizing AMPLILINK software v3.2 series (AL v3.2) there is a potential for wrong results to be assigned to an order.

Distribution pattern

Nationwide (USA) Distribution

Code information

05666678001