Z-2484-2020 Class II Terminated

Recalled by Ortho Clinical Diagnostics — Rochester, NY

Recall Details

Product Type
Devices
Report Date
July 8, 2020
Initiation Date
May 20, 2020
Termination Date
November 16, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total= 353 ( U.S.= 167; OUS 186)

Product Description

VITROS XT 7600 Integrated System Software Versions 3.5.1 and below Product Code:6844461 Unique Identifier: 10758750031610

Reason for Recall

Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances

Distribution Pattern

US Nationwide Distribution Foreign: Country AR AU BR CA CL CN CO DO FR ID IN JP KR MX MY NP PE PH SG TH TT VE VN

Code Information

Affected Serial Number Range: J76000111-J76000525