Z-2508-2023 Class I Ongoing

Recalled by Abbott Medical — Plano, TX

FDA device recall Z-2508-2023 was initiated by Abbott Medical on Jul 18, 2023 and is designated Class I. Reason for recall: Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators). The recall status is ongoing. Affected quantity: 387 devices (US only).

Recall details

Product Type
Devices
Report Date
Sep 20, 2023
Initiation Date
Jul 18, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
387 devices (US only)

Product description

Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).

Reason for recall

Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

Distribution pattern

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Code information

Model Number: 3672 UDI-DI Codes: 05415067046406 Serial Numbers: US CBW328.1 CCA167.1 CCA177.1 CCU613.1 CCV760.1 CCX700.1 CCY890.1 CDR307.1 CDW220.1 CEC548.1 CEC719.1 CEE095.1 CEH111.1 CEH156.1 CEH200.1 CEH272.1 CEH293.1 CEH334.1 CEH964.1 CEH975.1 CEH993.1 CEK127.1 CEK258.1 CEK265.1 CEK271.1 CEK310.1 CEK432.1 CEM232.1 CEM340.1 CEM349.1 CEM375.1 CEM383.1 CEM396.1 CEN429.1 CEN444.1 CEP494.1 CEP497.1 CEP537.1 CEP540.1 CEP541.1 CEP553.1 CEP554.1 CEP562.1 CEP585.1 CEP645.1 CEP658.1 CEP663.1 CEP671.1 CEP685.1 CEP725.1 CEP730.1 CEP838.1 CEP847.1 CEP867.1 CER393.1 CER395.1 CER462.1 CER482.1 CER509.1 CER576.1 CET134.1 CET147.1 CET298.1 CET336.1 CET343.1 CET649.1 CEY608.1 CEY648.1 CFL299.1 CFL403.1 CFL540.1 CFL582.1 CFP963.1 CFP987.1 CFP991.1 CFR079.1 CFR100.1 CFR121.1 CFR392.1 CFR446.1 CFR478.1 CFR585.1 CFU079.1 CFU082.1 CFU091.1 CFU097.1 CFU098.1 CFU099.1 CFU121.1 CFU126.1 CFU132.1 CFU145.1 CFU333.1 CFU401.1 CFU426.1 CFU441.1 CFU460.1 CFU467.1 CFU472.1 CFU483.1 CFU526.1 CFU547.1 CFU569.1 CFU646.1 CFU687.1 CFV457.1 CFV588.1 CFV718.1 CFX211.1 CGA013.1 CGA029.1 CGA065.1 CGA097.1 CGA145.1 CGA331.1 CGA365.1 CGA406.1 CGA426.1 CGA451.1 CGA491.1 CGA500.1 CGA515.1 CGA581.1 CGC898.1 CGC908.1 CGC916.1 CGC921.1 CGC926.1 CGC955.1 CGC957.1 CGD063.1 CGD070.1 CGD076.1 CGD085.1 CGD101.1 CGD111.1 CGD119.1 CGD123.1 CGD192.1 CGD193.1 CGD220.1 CGD231.1 CGD240.1 CGD248.1 CGD251.1 CGD277.1 CGD278.1 CGD283.1 CGD308.1 CGD342.1 CGD343.1 CGD362.1 CGD694.1 CGD719.1 CGD739.1 CGD777.1 CGD790.1 CGD832.1 CGD892.1 CGG890.1 CGH131.1 CGH175.1 CGH183.1 CGH199.1 CGH245.1 CGH346.1 CGH413.1 CGL108.1 CGL207.1 CGL321.1 CGL863.1 CGL866.1 CGL889.1 CGP051.1 CGP078.1 CGP326.1 CGP615.1 CGS914.1 CGW231.1 CHB011.1 CHE598.1 CHE845.1 CHF161.1 CHF178.1 CHH717.1 CHH744.1 CHH773.1 CHJ032.1 CHK331.1 CHK354.1 CHK434.1 CHK648.1 CHM211.1 CHM389.1 CHM509.1 CHP603.1 CHP605.1 CHP621.1 CHR155.1 CHR240.1 CHT217.1 CHT238.1 CHT240.1 CHT241.1 CHT243.1 CHT245.1 CHT272.1 CHT356.1 CHT361.1 CHT373.1 CHT469.1 CHT478.1 CHT539.1 CHT553.1 CHT561.1 CHT589.1 CHT609.1 CHT624.1 CHT758.1 CHT815.1 CHT834.1 CHV674.1 CHV680.1 CHV710.1 CHV713.1 CHV794.1 CHV901.1 CHV989.1 CHW217.1 CHW262.1 CHW263.1 CHW306.1 CHW350.1 CHW362.1 CHW526.1 CHW560.1 CHW590.1 CHW591.1 CHW640.1 CHZ471.1 CHZ563.1 CHZ576.1 CJA246.1 CJG129.1 CJG177.1 CJM932.1 CJM940.1 CJN773.1 CJN780.1 CJN879.1 CJN987.1 CJP047.1 CJW418.1 CJW475.1 CJW534.1 CJW548.1 CJW586.1 CJX925.1 CJX962.1 CJX982.1 CJX986.1 CJY012.1 CJY075.1 CJY111.1 CJY226.1 CJY238.1 CJY304.1 CJY323.1 CJY360.1 CJY409.1 CJZ061.1 CJZ103.1 CJZ135.1 CJZ157.1 CJZ224.1 CJZ241.1 CJZ253.1 CJZ256.1 CJZ283.1 CJZ293.1 CJZ300.1 CJZ330.1 CJZ420.1 CJZ444.1 CJZ450.1 CKB402.1 CKB428.1 CKB439.1 CKB447.1 CKB487.1 CKB507.1 CKB608.1 CKB624.1 CKB654.1 CKB696.1 CKB725.1 CKB767.1 CKB887.1 CKD072.1 CKF569.1 CKF587.1 CKG394.1 CKG550.1 CKG586.1 CKG607.1 CKG915.1 CKG954.1 CKG956.1 CKG962.1 CKH467.1 CKH606.1 CKH692.1 CKH719.1 CKH736.1 CKH765.1 CKH774.1 CKH786.1 CKH815.1 CKH859.1 CKH923.1 CKL195.1 CKL217.1 CKL236.1 CKL295.1 CKL345.1 CKL464.1 CKL487.1 CKP251.1 CKP255.1 CKP277.1 CKP291.1 CKP393.1 CKP435.1 CKP483.1 CKP495.1 CKP563.1 CKP656.1 CKP678.1 CKP682.1 CKP717.1 CKP733.1 CKP803.1 CKP939.1 CKS483.1 CKS519.1 CKS531.1 CKS552.1 CKS588.1 CKS590.1 CKS616.1 CKS641.1 CKS662.1 CKS673.1 CKS684.1 CKS692.1 CKS698.1 CKS706.1 CKS827.1 CKS856.1 CKS869.1 CKS954.1 CKS986.1 CKS992.1 CKS994.1 CKS997.1 CKT159.1 CKY723.1 CKY781.1 CLC604.1 CLD150.1 CLD293.1 CLF739.1 CLF855.1 CLF880.1 CLF888.1 CLF901.1 CLG005.1 CLG094.1 CLG124.1 CLG391.1 CLG393.1 CLG395.1 CLG396.1 CLG423.1 CLG457.1 CLG534.1 CLK164.1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2508-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Medical? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Abbott Medical: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Abbott Medical →

Source: openFDA record for Z-2508-2023. Report a data issue: hello@regdatalab.com.