Z-2555-2023 Class II
FDA device recall Z-2555-2023 was initiated by Landauer on Jul 12, 2023 and is designated Class II. Reason for recall: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential… Affected quantity: 16820 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 20, 2023
- Initiation Date
- Jul 12, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16820 units
Product description
nanoDot B - item 2 of 3 for 03500-000, Model Number 03502-000; radiation monitoring dosimeter used with the microSTAR readers
Reason for recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Distribution pattern
Worldwide
Code information
UDI/DI 0860003399903, all batch numbers