Z-2578-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-2578-2018 was initiated by GE Healthcare, LLC on May 24, 2018 and is designated Class II. Reason for recall: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). The recall status is terminated (terminated Feb 8, 2019). Affected quantity: 42 devices total.

Recall Details

Product Type
Devices
Report Date
Aug 8, 2018
Initiation Date
May 24, 2018
Termination Date
Feb 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 devices total

Product Description

Discovery IQ 2.0 20cm, 4 ring, Model Number 5432539-22

Reason for Recall

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Distribution Pattern

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Code Information

Model Number 5432539-22. Serial Number 502674HM5 (System ID Number 510204DISCIQ); Serial Number 504305HM4 (System ID Number 219922PTIQ)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2578-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.