Z-2788-2016 Class II Terminated

Recalled by Howmedica Osteonics Corp. — Allendale, NJ

FDA device recall Z-2788-2016 was initiated by Howmedica Osteonics Corp. on Jul 26, 2016 and is designated Class II. Reason for recall: Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence… The recall status is terminated (terminated Mar 12, 2018).

Recall details

Product Type
Devices
Report Date
Sep 21, 2016
Initiation Date
Jul 26, 2016
Termination Date
Mar 12, 2018
Voluntary/Mandated
Voluntary: Firm initiated

Product description

AccuLIF PL 10-16mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the AccuLIF PL Cages can be used as in adjunct to fusion in patients diagnosed with degenerative scoliosis. The AccuLIF Pl Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF PL Cages are to be used with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion

Reason for recall

Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported post-operative reduction of cage height (> 1mm) of AccuLIF PL Expandable Cages is approximately 0.47% and has demonstrated a slight upward trend since September 2015.

Distribution pattern

NationwideDistribution

Code information

Catalog #400010 Lot #'s 01021403, 01121508, 01121509, 01271515, 01271516, 02021512, 02021513, 02021514, 02021515, 02021602, 03221310, 04141508, 04141509, 04141510, 04141511, 04141512, 04251312, 06101405, 06101406, 06101407, 06251311, 07011406, 07201510, 07201511, 07201512, 08141303, 08141313, 09251504, 09281512, 10081301, 10081302, 10201502, 10201503, 10301504, 11121514, 11121518, 11121521, 12151511 and 12151517

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2788-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2788-2016. Report a data issue: hello@regdatalab.com.