Z-2903-2026 Class II Ongoing
FDA device recall Z-2903-2026 was initiated by TISSIUM SA on Jun 15, 2026 and is designated Class II. Reason for recall: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated… The recall status is ongoing.
Recall details
- Product Type
- Devices
- Report Date
- Aug 12, 2026
- Initiation Date
- Jun 15, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Reason for recall
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Distribution pattern
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Code information
Lot Code: Model No NCOCO2010; UDI-DI 03760279380094; Lot Number 250700042, 251000033, and 251000034