Z-2906-2026 Class II Ongoing

Recalled by TISSIUM SA — Paris 12e Arrondissement

FDA device recall Z-2906-2026 was initiated by TISSIUM SA on Jun 15, 2026 and is designated Class II. Reason for recall: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Aug 12, 2026
Initiation Date
Jun 15, 2026
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Reason for recall

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Distribution pattern

US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.

Code information

Lot Code: Model No NCOCO4010; UDI-DI 03760279380148; Lot Numbers 251000041, 250700048, 260300046, 260400050, and 260300047

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2906-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2906-2026. Report a data issue: hello@regdatalab.com.