Z-2905-2026 Class II Ongoing

Recalled by TISSIUM SA — Paris 12e Arrondissement

FDA device recall Z-2905-2026 was initiated by TISSIUM SA on Jun 15, 2026 and is designated Class II. Reason for recall: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Aug 12, 2026
Initiation Date
Jun 15, 2026
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Reason for recall

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Distribution pattern

US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.

Code information

Lot Code: Model No NCOCO3010; UDI-DI 03760279380117; Lot Numbers 250700045, 251000040, 251000036, 251000037, 251000039, 251000035, and 251000064

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2905-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2905-2026. Report a data issue: hello@regdatalab.com.