Z-2936-2018 Class II Terminated

Recalled by Medical Components, Inc Dba MedComp — Harleysville, PA

FDA device recall Z-2936-2018 was initiated by Medical Components, Inc Dba MedComp on Jul 12, 2018 and is designated Class II. Reason for recall: The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings)… The recall status is terminated (terminated Jul 2, 2020).

Recall details

Product Type
Devices
Report Date
Sep 5, 2018
Initiation Date
Jul 12, 2018
Termination Date
Jul 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Duo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits

Reason for recall

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

Distribution pattern

Nationwide distribution; worldwide distribution.

Code information

MKAZ210 MCBK870 MCBS910 MCBZ450 MCCP990 MCCZ440 MCDNlOO MCDQ930 MCFD710 MCFN970 MCFS730 MCAS340 MCBA830 MCBR880 MCBY910 MCCF130 MCCX150 MCDF010 MCDG980 MCDP120 MCFD720 MCFJ060 MCFM520 MCFW300 MCFZ590 MKAJ380 MCAS230 MCBL150 MCBN500 MCBZ380 MCCV010 MCDF030 MCFX080 MCFV620 MCBA170 MCDC940 MKAX240 MCAC870 MCCQ000 MCFG380 MKAQ920 MCAH920 MCCW950

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2936-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medical Components, Inc Dba MedComp? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medical Components, Inc Dba MedComp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medical Components, Inc Dba MedComp →

Source: openFDA record for Z-2936-2018. Report a data issue: hello@regdatalab.com.