Z-2938-2018 Class II Terminated

Recalled by Medical Components, Inc Dba MedComp — Harleysville, PA

FDA device recall Z-2938-2018 was initiated by Medical Components, Inc Dba MedComp on Jul 12, 2018 and is designated Class II. Reason for recall: The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings)… The recall status is terminated (terminated Jul 2, 2020).

Recall details

Product Type
Devices
Report Date
Sep 5, 2018
Initiation Date
Jul 12, 2018
Termination Date
Jul 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Duo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M

Reason for recall

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

Distribution pattern

Nationwide distribution; worldwide distribution.

Code information

MHJV260 MHMJ380 MHRC840 MHXF770 MJAZ910 MJDV230 MJLF440 MLCT530 MCAH650 MCAQ240 MCAW220 MCBC440 MCCJ920 MCCP260 MCDC910 MCDG320 MCDL590 MCDW940 MCFA540 MCFG000 MCFM820 MCFR060 MCFV060 MKAA520 MKAL360 MKAN510 MKAQ280 MKAZ530 MCAA590 MCAD760 MCAK990 MCBM860 MCBQ500 MCCA150 MCCG660 MCDD810 MCDG250 MCDJ980 MCDP910 MCDV760 MKAD560 MKAL370 MKBC340 MBZS180 MCAH050 MCAL000 MCBM870 MCBT030 MCCG580 MCCP590 MCDD570 MCDR280 MCDX060 MCFQ250 MCFS300 MCFT650 MCFX880 MKAB600 MKAM870 MBZY670 MCCT820 MCFQ150 MKBB070 MJLP070 MLDH060 MHLM190 MHRX850 MHSY930 MHYG710 MJCV440 MBYR870 MBZM650 MCAH670 MCAJ630 MCAQ250 MCAQ870 MCBA060 MCBD890 MCBX380 MCDF830 MCDL020 MCDR290 MCDX070 MCFA570 MCFD860 MCFH790 MKAF320 MKAG530 MKAK390 MKAL400 MKAM880 MKBA670 MCCT830 MHLJ070 MHRD620 MHRX860 MJKJ070 MCBC830 MCFK400

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2938-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medical Components, Inc Dba MedComp? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medical Components, Inc Dba MedComp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medical Components, Inc Dba MedComp →

Source: openFDA record for Z-2938-2018. Report a data issue: hello@regdatalab.com.