Z-0123-2017 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-0123-2017 was initiated by Randox Laboratories Ltd. on Aug 2, 2016 and is designated Class II. Reason for recall: According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all… The recall status is terminated (terminated Sep 7, 2017). Affected quantity: 18.

Recall details

Product Type
Devices
Report Date
Oct 26, 2016
Initiation Date
Aug 2, 2016
Termination Date
Sep 7, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18

Product description

Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.

Reason for recall

According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.

Distribution pattern

US Nationwide Distribution

Code information

Catalogue number: LI3837, LI7979, LI8050, LI8361

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0123-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Randox Laboratories Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0123-2017. Report a data issue: hello@regdatalab.com.