Z-0123-2017 Class II Terminated
FDA device recall Z-0123-2017 was initiated by Randox Laboratories Ltd. on Aug 2, 2016 and is designated Class II. Reason for recall: According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all… The recall status is terminated (terminated Sep 7, 2017). Affected quantity: 18.
Recall details
- Product Type
- Devices
- Report Date
- Oct 26, 2016
- Initiation Date
- Aug 2, 2016
- Termination Date
- Sep 7, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 18
Product description
Reason for recall
Distribution pattern
US Nationwide Distribution
Code information