Z-0213-2026 Class II Ongoing

Recalled by Wipro GE Healthcare Private Ltd. — Bengaluru

FDA device recall Z-0213-2026 was initiated by Wipro GE Healthcare Private Ltd. on September 5, 2025 and is designated Class II. Reason for recall: Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. Thi… The recall status is ongoing. Affected quantity: 173 units.

Recall Details

Product Type
Devices
Report Date
October 29, 2025
Initiation Date
September 5, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
173 units

Product Description

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

Reason for Recall

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Distribution Pattern

Worldwide distribution.

Code Information

GTIN 00195278343512, Serial Numbers: STU25210418SA, STU25210417SA, STU25210419SA, STU25250440SA, STU25250439SA, STU25260447SA, STU25210407SA, STU25210406SA, STU24200003SA, STU24200006SA, STU24200005SA, STU24200004SA, STU23240015SA, STU24240019SA, STU24240020SA, STU24240007SA, STU24240022SA, STU24240021SA, STU24330020SA, STU24240014SA, STU24240013SA, STU24330019SA, STU24330022SA, STU24330016SA, STU24240005SA, STU24240008SA, STU24240012SA, STU24330018SA, STU24330021SA, STU24330017SA, STU23160011SA, STU23240007SA, STU24310001SA, STU25210433SA, STU25210432SA, STU25200390SA, STU25200392SA, STU25200391SA, STU25200399SA, STU25200397SA, STU25200395SA, STU25200393SA, STU25200398SA, STU25200396SA, STU25250438SA, STU25270475SA, STU25270456SA, STU25270468SA, STU25270458SA, STU25270466SA, STU25270476SA, STU25270469SA, STU25270473SA, STU25270459SA, STU25270474SA, STU25270467SA, STU25210430SA, STU25210429SA, STU25180343SA, STU25190347SA, STU25190348SA, STU24200007SA, STU25190360SA, STU25240436SA, STU25190361SA, STU25190362SA, STU25190363SA, STU25240437SA, STU25190359SA, STU24510027SA, STU24510032SA, STU24510010SA, STU24510020SA, STU24510017SA, STU24510026SA, STU24510018SA, STU24510024SA, STU24510025SA, STU24510031SA, STU24510011SA, STU24510029SA, STU24510023SA, STU24510030SA, STU24510021SA, STU24510019SA, STU24510022SA, STU25190350SA, STU25200388SA, STU25200389SA, STU25210412SA, STU25190349SA, STU25190381SA, STU25190376SA, STU25190383SA, STU25190374SA, STU25190367SA, STU25190372SA, STU25190379SA, STU25190368SA, STU25190352SA, STU25190358SA, STU25190370SA, STU25190366SA, STU25190355SA, STU25190354SA, STU25190382SA, STU25190380SA, STU25190371SA, STU25190377SA, STU25190353SA, STU25190351SA, STU25190364SA, STU25190357SA, STU25190384SA, STU25190369SA, STU25190373SA, STU25190356SA, STU25190365SA, STU25190378SA, STU25190385SA, STU25210410SA, STU25210411SA, STU25210409SA, STU25230435SA, STU25180346SA, STU25180344SA, STU25180345SA, STU25210413SA, STU25210414SA, STU25210423SA, STU25210431SA, STU25210415SA, STU25270453SA, STU25260450SA, STU25260451SA, STU25290488SA, STU25210408SA, STU25210400SA, STU25270454SA, STU25210404SA, STU25210421SA, STU25270464SA, STU25210420SA, STU25270462SA, STU25210425SA, STU25270461SA, STU25210426SA, STU25210428SA, STU25210427SA, STU25270463SA, STU25270460SA, STU24460058SA, STU24460059SA, STU24460071SA, STU24460060SA, STU25200387SA, STU25210401SA, STU25200386SA, STU25210424SA, STU25270457SA, STU25270465SA, STU25300492SA, STU25210402SA, STU25210403SA, STU25250446SA, STU25210422SA, STU25210405SA, STU25260449SA, STU25290481SA, STU25290482SA, STU25290483SA, STU25290484SA, STU25210416SA.