Z-0958-2024 Class I Ongoing

Recalled by Wipro GE Healthcare Private Ltd. — Bengaluru, N/A

FDA device recall Z-0958-2024 was initiated by Wipro GE Healthcare Private Ltd. on Dec 26, 2023 and is designated Class I. Reason for recall: GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be… The recall status is ongoing. Affected quantity: 1410 units.

Recall details

Product Type
Devices
Report Date
Feb 21, 2024
Initiation Date
Dec 26, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1410 units

Product description

GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR

Reason for recall

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Distribution pattern

Worldwide distribution.

Code information

a) Lullaby Incubator, Model Numbers: M1151042, QCR031; b) Lullaby Incubator XP, Model Numbers: 8320, M1151135, QCR034; c) Lullaby Incubator TR, Model Numbers: M1151357 ALL SERIAL NUMBERS

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0958-2024
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0958-2024. Report a data issue: hello@regdatalab.com.