Z-0961-2024 Class I Ongoing
FDA device recall Z-0961-2024 was initiated by Wipro GE Healthcare Private Ltd. on Dec 26, 2023 and is designated Class I. Reason for recall: GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be… The recall status is ongoing. Affected quantity: 1 unit.
Recall details
- Product Type
- Devices
- Report Date
- Feb 21, 2024
- Initiation Date
- Dec 26, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product description
GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836, M1166814
Reason for recall
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
Distribution pattern
Worldwide distribution.
Code information
ALL SERIAL NUMBERS