Z-0959-2024 Class I Ongoing

Recalled by Wipro GE Healthcare Private Ltd. — Bengaluru, N/A

Recall Details

Product Type
Devices
Report Date
February 21, 2024
Initiation Date
December 26, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1465 units

Product Description

GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850

Reason for Recall

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Distribution Pattern

Worldwide distribution.

Code Information

ALL SERIAL NUMBERS