Z-0282-2020 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-0282-2020 was initiated by Randox Laboratories Ltd. on Aug 9, 2019 and is designated Class II. Reason for recall: The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to… The recall status is terminated (terminated Jan 8, 2021). Affected quantity: 17 kits.

Recall details

Product Type
Devices
Report Date
Nov 13, 2019
Initiation Date
Aug 9, 2019
Termination Date
Jan 8, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17 kits

Product description

Rheumatoid Factor Standard (RF CAL) Catalogue Number: RF2301 GTIN: 05055273205039 Product Usage: The Randox Rheumatoid Factor Latex-Enhanced Immunorutbidimetric Test Kit is an in vitro diagnostic reagent for the quantitative determination of rheumatoid factors (RF) in human serum. This test kit may be used to facilitate the detection and diagnosis of Rheumatoid Arthritis. RF have also been observed in the serum of patients with systemic lupus eruthematosus, hepatitis. Liver cirrhosis and syphilis. However, in these conditions the RF titre is much lower than in rheumatoid arthritis. This test kit must be used by suitably qualified personnel under appropriate laboratory conditions. RF calibrator and RF positive control are for use in the calibration and quality control of RF latex-enhanced immunoturbidimetric assays.

Reason for recall

The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.

Distribution pattern

US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.

Code information

Batch/Lot # 463479/1325 -1329RF 466715/1325 -1329RF 474059/1339 1343RF 480410/ 1347 1351RF

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0282-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Randox Laboratories Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0282-2020. Report a data issue: hello@regdatalab.com.