FDA device recall Z-0283-2020 was initiated by Randox Laboratories Ltd. on August 9, 2019 and is designated Class II. Reason for recall: The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to t… The recall status is terminated (terminated January 8, 2021). Affected quantity: 100 kits.
Recall Details
- Product Type
- Devices
- Report Date
- November 13, 2019
- Initiation Date
- August 9, 2019
- Termination Date
- January 8, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 100 kits
Product Description
Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems
Reason for Recall
The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.
Distribution Pattern
US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.
Code Information
Batch/Lot # 440973 /1748 1752IT 451414 /1748 1752IT 470531 /1813 1817IT 480859 /1813 1817IT