Z-0295-2017 Class II Terminated

Recalled by Ethicon Endo-Surgery Inc — Blue Ash, OH

FDA device recall Z-0295-2017 was initiated by Ethicon Endo-Surgery Inc on Jul 15, 2016 and is designated Class II. Reason for recall: Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device… The recall status is terminated (terminated Feb 28, 2017). Affected quantity: 105.

Recall details

Product Type
Devices
Report Date
Oct 26, 2016
Initiation Date
Jul 15, 2016
Termination Date
Feb 28, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
105

Product description

Endoscopic Articulating Linear Cutter, ATB35. Standard 35 MM. The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses

Reason for recall

Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.

Distribution pattern

Worldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.

Code information

N91L48, N91L49, N91R40, N91R41, N91U0W, N91U0X

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0295-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ethicon Endo-Surgery Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ethicon Endo-Surgery Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ethicon Endo-Surgery Inc →

Source: openFDA record for Z-0295-2017. Report a data issue: hello@regdatalab.com.