Z-0328-2022 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North)

FDA device recall Z-0328-2022 was initiated by Randox Laboratories Ltd. on Nov 1, 2021 and is designated Class II. Reason for recall: When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range The recall status is terminated (terminated Mar 9, 2023). Affected quantity: 15 kits.

Recall details

Product Type
Devices
Report Date
Dec 8, 2021
Initiation Date
Nov 1, 2021
Termination Date
Mar 9, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 kits

Product description

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Reason for recall

When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Distribution pattern

US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.

Code information

Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0328-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Randox Laboratories Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0328-2022. Report a data issue: hello@regdatalab.com.