Z-0547-2021 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
December 16, 2020
Initiation Date
November 13, 2020
Termination Date
April 1, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 kits

Product Description

Randox Rx Series Mg - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG8326

Reason for Recall

Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine

Distribution Pattern

US Nationwide distribution in the states of GA, NJ.

Code Information

Lot Number: 539770 Exp. Date : 28 March 2022 UDI: 05055273209303