Z-0577-2015 Class II Terminated
FDA device recall Z-0577-2015 was initiated by Invacare Corporation on Sep 23, 2014 and is designated Class II. Reason for recall: It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen… The recall status is terminated (terminated Jun 18, 2015). Affected quantity: 5,963.
Recall details
- Product Type
- Devices
- Report Date
- Dec 17, 2014
- Initiation Date
- Sep 23, 2014
- Termination Date
- Jun 18, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,963
Product description
Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
Reason for recall
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
Distribution pattern
Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
Code information
The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)