Z-0718-2020 Class II Terminated

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

FDA device recall Z-0718-2020 was initiated by Bard Peripheral Vascular Inc on Feb 18, 2019 and is designated Class II. Reason for recall: Certain product code/lot number combinations may experience introducer sheath compatibility issues when used with a 6F introducer sheath. Using the affected balloon expandable vascular covered stents… The recall status is terminated (terminated Apr 28, 2021). Affected quantity: 127 Units.

Recall details

Product Type
Devices
Report Date
Dec 25, 2019
Initiation Date
Feb 18, 2019
Termination Date
Apr 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
127 Units

Product description

LIFESTREAM Balloon Expandable Vascular Covered Stent . BARD PERIPHERAL VASCULAR. 6mm X 26mm X 80 Recommended Introducer 6F (2.00mm) NP 8atm (811 kPa) Recommended Guide wire .035" (0.89mm) RBP 12 atm (1216 kPa) RX Only Lot numbers begins with CMA or CMB. Product Code Stent Diameter Stent length Catheter Length LSM0800626 6mm 26 mm 80 cm LSM1350626 6 mm 26 mm 135 cm LSM0800637 6 mm 37 mm 80 cm LSM1350637 6 mm 37 mm 135 cm LSM0800726 7 mm 26 mm 80 cm LSM1350726 7 mm 26 mm 135 cm Product Usage: The LIFESTREAM Balloon Expandable Vascular Covered Stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries with reference vessel diameters between 4.5 mm and 12.0 mm, and lesions lengths up to 100mm.

Reason for recall

Certain product code/lot number combinations may experience introducer sheath compatibility issues when used with a 6F introducer sheath. Using the affected balloon expandable vascular covered stents with a 6F introducer sheath could result in difficulty or the inability to insert or advance the catheter through the introducer sheath. This could potentially lead to a slight prolongation of the procedure.

Distribution pattern

127 units were distributed in Australia to 32 Consignees.

Code information

Begins with Lot Number CMA or CMB Product Code LSM0800626 Product Code LSM1350626 Product Code LSM0800637 Product Code LSM1350637 Product Code LSM0800726 Product Code LSM1350726

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0718-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bard Peripheral Vascular Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Bard Peripheral Vascular Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Bard Peripheral Vascular Inc →

Source: openFDA record for Z-0718-2020. Report a data issue: hello@regdatalab.com.