Z-0762-2017 Class II Terminated

Recalled by Randox Laboratories — Antrim

FDA device recall Z-0762-2017 was initiated by Randox Laboratories on Sep 16, 2016 and is designated Class II. Reason for recall: An internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for… The recall status is terminated (terminated Apr 11, 2018). Affected quantity: 2817.

Recall details

Product Type
Devices
Report Date
Dec 21, 2016
Initiation Date
Sep 16, 2016
Termination Date
Apr 11, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2817

Product description

The Randox Liquid Cardiac Controls Level 1, Level 2, and Level 3 are liquid controls containing BNP, CK MB Mass, Digoxin, D-Dimer, Homocysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I, and Troponin T. They have been developed for use in the quality control of BNP, CK MB Mass, Digoxin, D-Dimer, Homocysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I, and Troponin T assays on various clinical chemistry and immunoassay systems. This in vitro diagnostic device is intended for prescription use only.

Reason for recall

An internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for Roche e411, BioMerieux Vidas, and Siemens Centaur Classic methods.

Distribution pattern

Worldwide Distribution - US Distribution.

Code information

CQ5053 CQ5052

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0762-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Randox Laboratories? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0762-2017. Report a data issue: hello@regdatalab.com.