Z-0911-2016 Class I Terminated

Recalled by Abbott Vascular — Temecula, CA

FDA device recall Z-0911-2016 was initiated by Abbott Vascular on Feb 4, 2016 and is designated Class I. Reason for recall: Abbott Vascular has recently received reports of cases on Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted implanting a MitraClip, but the Clip could not be… The recall status is terminated (terminated May 12, 2016). Affected quantity: 7,600 units total (2,300 units in US).

Recall details

Product Type
Devices
Report Date
Mar 16, 2016
Initiation Date
Feb 4, 2016
Termination Date
May 12, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,600 units total (2,300 units in US)

Product description

MitraClip Clip Delivery System, product number MSK0101. The MitraClip System contains the Clip Delivery System, product number CDS0201, GTIN 08717648195914 and the Steerable Guide Catheter, product number SGC0101, GTIN 08717648195921. The MitraClip Clip Delivery System (CDS) consists of three major components: 1) the Delivery Catheter 2) the Steerable Sleeve, and 3) the MitraClip device. The implantable MitraClip device is located at the distal end of the CDS. The CDS is used to advance and manipulate the implantable MitraClip device for proper positioning and placement on the mitral valve leaflets. The Delivery Catheter, which is part of the CDS, is designed to deliver and deploy the MitraClip. The Delivery Catheter extends through the thru-lumen of the Steerable Sleeve.

Reason for recall

Abbott Vascular has recently received reports of cases on Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted implanting a MitraClip, but the Clip could not be detached from the delivery system due to a mandrel fracture.

Distribution pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Argentina Luxembourg Australia Malaysia Austria Mexico Belgium Netherlands Brazil New Zealand Canada Norway Colombia Philippines Corrientes Poland Czech Republic Portugal Denmark Egypt Romania Finland Saudi Arabia France Slovenia Germany Spain Greece Sweden Hong Kong Switzerland Hungary Thailand Indonesia Turkey Israel United Arab Emirates Italy United Kingdom Kuwait and Vietnam.

Code information

Lot number 50811U1 50811U2 50812U1 50813U1 50814U1 50826U1 50826U2 50827U1 50908U1 50908U2 50909U1 50910U1 50910U2 50911U1 50924U1 50925U1 50928U1 50929U1 51010U1 51012U1 51012U2 51013U1 51014U1 51014U2 51026U1 51027U3 51028U1 51028U2 51029U1 51030U1 51105U1 51106U1 51109U1 51109U2 51110U1 51110U2 51117U1 51203U1 51204U1 51205U1 51207U1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0911-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Vascular? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0911-2016. Report a data issue: hello@regdatalab.com.