Z-0917-2020 Class II Terminated

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

FDA device recall Z-0917-2020 was initiated by Bard Peripheral Vascular Inc on Aug 14, 2019 and is designated Class II. Reason for recall: The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion… The recall status is terminated (terminated Feb 26, 2021). Affected quantity: 286 Units.

Recall details

Product Type
Devices
Report Date
Feb 5, 2020
Initiation Date
Aug 14, 2019
Termination Date
Feb 26, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
286 Units

Product description

Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F; Nominal Pressure 8atm, Rated Burst Pressure 40atm; Syringe inflation 1cc; UDI: (01)00801741060533(17)220228(10)REDQ3444

Reason for recall

The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists that upon inflation, non-diseased vessel, kidney or sections of the AV dialysis fistulae may be dilated, potentially causing a varying degree of injury. This may require additional medical or surgical intervention.

Distribution pattern

US: NJ, NY, OH, PA, VA, SC, GA, AR, TX, MA, NC, TN, LA, AZ, MS, FL, AL,MD, CA, HI, MO, CO, WI, DC, KY, IL, MN, IN, KS, OK, CT, MT, WA, ND, IA, WA, WV, PR OUS: None

Code information

Lot Number REDQ3444 Exp: 02/28/2022

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0917-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bard Peripheral Vascular Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Bard Peripheral Vascular Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Bard Peripheral Vascular Inc →

Source: openFDA record for Z-0917-2020. Report a data issue: hello@regdatalab.com.