Z-1100-2020 Class III Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
February 12, 2020
Initiation Date
December 13, 2019
Termination Date
October 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product Description

Liquid Clinical Chemistry Control, Catalogue Number: LAE4217, Size: 12 x 5 ml

Reason for Recall

A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86).

Distribution Pattern

US: NY and DE OUS: France, Lithuania, Poland, Spain, Switzerland, Turkey, Australia, Hong Kong, Bosnia and Herzegovina, Costa Rica, Ecuador, Gambia, New Zealand and Solomon Islands.

Code Information

Catalogue Number: LAE4217 GTIN: 05055273208993 Lot Number: 1021UE (EXP: 2020-06-28)