Z-1126-2013 Class II Terminated

Recalled by Accuray Incorporated — Sunnyvale, CA

Recall Details

Product Type
Devices
Report Date
April 24, 2013
Initiation Date
March 19, 2013
Termination Date
October 23, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
176 units

Product Description

Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment.

Reason for Recall

A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.

Distribution Pattern

Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.

Code Information

All units of these models.