Z-1530-2013 Class II Terminated
Recalled by Ethicon Endo-Surgery Inc
— Cincinnati, OH
FDA device recall Z-1530-2013 was initiated by Ethicon Endo-Surgery Inc on May 13, 2013 and is designated Class II. Reason for recall: Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 60mm Black Reload (ECR60T) due to the potential for incomplete staple line formation from reload damage during the firing… The recall status is terminated (terminated Feb 11, 2014). Affected quantity: 57,540 Units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 19, 2013
- Initiation Date
- May 13, 2013
- Termination Date
- Feb 11, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 57,540 Units
Product description
ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON and ECHELON FLEX families of Endoscopic Linear Cutters (articulating and straight) are sterile, single patient use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON 60 instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The shaft can rotate freely in both directions and an articulation mechanism on articulating instruments enables bending the distal portion of the shaft to facilitate lateral access of the operative site. The instruments are shipped without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments'¿" lock-out feature is designed to prevent a used reload from being refired.
Reason for recall
Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 60mm Black Reload (ECR60T) due to the potential for incomplete staple line formation from reload damage during the firing sequence.
Distribution pattern
Worldwide Distribution - US Nationwide all fifty states and including the countries of Canada, Belgium, India, Malaysia, Saudi Arabia, Brazil, Israel, Mexico, Singapore, Jordan, New Zealand, Taiwan, Chile, Korea, Puerto Rico, Turkey, Columbia, Kuwait, Qatar, United Arab Emirates, Egypt, Lebanon and Russia.
Code information
The following lot numbers and associated expiration dates are affected by this recall: J4AY8Z 5/15/2013, J4C00J 6/15/2013, J4C87R 7/15/2013, J4CC1F 8/15/2013, J4CD2X 8/15/2013, J4CF3D 9/15/2013, J4CJ0E 9/15/2013, J4CJ78 9/15/2013, J4CK2P 9/15/2013, J4CM9X 10/15/2013, J4CR2H 10/15/2013, J4CT5H 11/15/2013, J4CU86 11/15/2013, J4CV1W 11/15/2013, J4CV5H 11/15/2013, K4C149 12/15/2013, K4C17W 12/15/2013, K4C33U 12/15/2013, K4C41C 12/15/2013, K4C600 1/15/2014, K4C64A 1/15/2014, 4C728 1/15/2014, K4C82K 1/15/2014, K4C88F 1/15/2014, K4C94W 1/15/2014, K4CC0L 2/15/2014, CC8L 2/15/2014, K4CD1E 2/15/2014, K4CF1P 2/15/2014, K4CF5D 2/15/2014, K4CG18 3/15/2014, K4CG4J 3/15/2014, K4CG8P 3/15/2014, K4CH44 3/15/2014, K4CK0W 3/15/2014,
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1530-2013
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Ethicon Endo-Surgery Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Ethicon Endo-Surgery Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Ethicon Endo-Surgery Inc →
Use it in ChatGPT
Get a free API key
How the connector works