Z-1547-2020 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-1547-2020 was initiated by Randox Laboratories Ltd. on February 18, 2020 and is designated Class II. Reason for recall: The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly. The recall status is ongoing. Affected quantity: 186 total, 27 USA.

Recall Details

Product Type
Devices
Report Date
April 1, 2020
Initiation Date
February 18, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
186 total, 27 USA

Product Description

Chemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 - Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The device is for the control of accuracy.

Reason for Recall

The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of WV, IN, VA, ID, PR, DE, and countries of West Indies, Czech Republic, France, Germany, Hong Kong, Ireland, Italy, Norway, Philippines, South Africa, South Korea, Sweden, Taiwan (R.O.C.), Thailand, UAE, UK.

Code Information

Lot 306UL