Z-2151-2017 Class I Terminated
Recalled by Abbott Vascular
— Temecula, CA
FDA device recall Z-2151-2017 was initiated by Abbott Vascular on Apr 26, 2017 and is designated Class I. Reason for recall: Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and… The recall status is terminated (terminated Nov 16, 2017). Affected quantity: 590 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 7, 2017
- Initiation Date
- Apr 26, 2017
- Termination Date
- Nov 16, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 590 units
Product description
Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.
Reason for recall
Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
Distribution pattern
Australia
Code information
Australia: Part No. Description 1012462-08 2.5 x 08 mm Absorb BVS 1012462-12 2.5 x 12 mm Absorb BVS 1012462-18 2.5 x 18 mm Absorb BVS 1012462-23 2.5 x 23 mm Absorb BVS 1012462-28 2.5 x 28 mm Absorb BVS 1012463-08 3.0 x 08 mm Absorb BVS 1012463-12 3.0 x 12 mm Absorb BVS 1012463-18 3.0 x 18 mm Absorb BVS 1012463-23 3.0 x 23 mm Absorb BVS 1012463-28 3.0 x 28 mm Absorb BVS 1012464-12 3.5 x 12 mm Absorb BVS 1012464-18 3.5 x 18 mm Absorb BVS 1012464-23 3.5 x 23 mm Absorb BVS 1012464-28 3.5 x 28 mm Absorb BVS
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2151-2017
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Abbott Vascular? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works