Z-2151-2017 Class I Terminated

Recalled by Abbott Vascular — Temecula, CA

Recall Details

Product Type
Devices
Report Date
June 7, 2017
Initiation Date
April 26, 2017
Termination Date
November 16, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
590 units

Product Description

Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.

Reason for Recall

Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.

Distribution Pattern

Australia

Code Information

Australia: Part No. Description 1012462-08 2.5 x 08 mm Absorb BVS 1012462-12 2.5 x 12 mm Absorb BVS 1012462-18 2.5 x 18 mm Absorb BVS 1012462-23 2.5 x 23 mm Absorb BVS 1012462-28 2.5 x 28 mm Absorb BVS 1012463-08 3.0 x 08 mm Absorb BVS 1012463-12 3.0 x 12 mm Absorb BVS 1012463-18 3.0 x 18 mm Absorb BVS 1012463-23 3.0 x 23 mm Absorb BVS 1012463-28 3.0 x 28 mm Absorb BVS 1012464-12 3.5 x 12 mm Absorb BVS 1012464-18 3.5 x 18 mm Absorb BVS 1012464-23 3.5 x 23 mm Absorb BVS 1012464-28 3.5 x 28 mm Absorb BVS