Z-2371-2019 Class II Terminated

Recalled by Bard Peripheral Vascular Inc — Tempe, AZ

Recall Details

Product Type
Devices
Report Date
September 4, 2019
Initiation Date
May 31, 2019
Termination Date
May 17, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1356 units

Product Description

EQUISTREAM Long-Term Hemodialysis Catheter with Preloaded Styet, REF/Product code:5903190, 14.5F, Straight ,19cm length, BARD, UDI# 00801741013690

Reason for Recall

The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.

Distribution Pattern

US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA,MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY OUS: Australia, Canada,

Code Information

(Lot #, Exp.) RECZ3009 8/31/2020 RECZ0716 7/31/2020 RECW0585 7/31/2020 RECX1187 8/31/2020 RECV1624 1/31/2020