Z-2374-2015 Class I Terminated
FDA device recall Z-2374-2015 was initiated by Ethicon Endo-Surgery Inc on May 13, 2015 and is designated Class I. Reason for recall: Ethicon has found that disinfecting methods not specified in the Instructions For Use (IFU) have led to the degradation of the plastic material of the SEDASYS System Control Unit. The recall status is terminated (terminated May 2, 2016). Affected quantity: 14 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 2, 2015
- Initiation Date
- May 13, 2015
- Termination Date
- May 2, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14 units
Product description
Reason for recall
Distribution pattern
Nationwide Distribution including OH, RI, TX & WA.