Z-2446-2023 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-2446-2023 was initiated by Randox Laboratories Ltd. on Jun 21, 2023 and is designated Class II. Reason for recall: (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error… The recall status is ongoing. Affected quantity: 39 units.

Recall details

Product Type
Devices
Report Date
Aug 30, 2023
Initiation Date
Jun 21, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
39 units

Product description

Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine, Glucose, hCG pregnancy, Magnesium, Microalbumin, Osmolality, pH, Phosphate Inorganic, Potassium, Protein Total, Sodium, Specific Gravity, Urea and Uric Acid (Urate) on clinical chemistry systems Catalog Number: UC5074

Reason for recall

(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AR, FL, GA, OH, VA including PR and WW distribution.

Code information

GTIN: 05055273207569 Batch/Lot Number: 1209UC Expiry Date: 28 Mar 24

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2446-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Randox Laboratories Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2446-2023. Report a data issue: hello@regdatalab.com.