Z-2448-2023 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-2448-2023 was initiated by Randox Laboratories Ltd. on Jul 19, 2023 and is designated Class II. Reason for recall: As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material… The recall status is ongoing. Affected quantity: 166 units.

Recall details

Product Type
Devices
Report Date
Aug 30, 2023
Initiation Date
Jul 19, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
166 units

Product description

Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691

Reason for recall

As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.

Distribution pattern

CA, FL, IL, ME, MI, NC, OH, PA,NJ,

Code information

GTIN: 05055273204032 Batch/ Lot number Expiry Date 590858 28 08 23; 590859 28 08 23; 627222 28Jul 24; 627224 28Jul24; 634886 28Jul24; 634887 28Jul24

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2448-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.