Z-2751-2020 Class II Terminated

Recalled by RANDOX LABORATORIES, LTD. — Crumlin Colorado Antrim

Recall Details

Product Type
Devices
Report Date
August 12, 2020
Initiation Date
June 26, 2020
Termination Date
October 19, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2219 kits

Product Description

Randox Lipase Assay: Lipase Colorimetric Reagent - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.

Reason for Recall

Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase reagent batches.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of Puerto Rico and WV.

Code Information

Catalogue Numbers: LI3837, LI7979, LI8050, LI8361 GTIN: 05055273204230, 05055273204247, 05055273209136, 05055273214284 Lot #: 507360, 497086, 497053, 497038, 502135