Z-2826-2018 Class I Terminated

Recalled by Vyaire Medical — Mettawa, IL

Recall Details

Product Type
Devices
Report Date
August 29, 2018
Initiation Date
June 20, 2018
Termination Date
July 28, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2150 Units Total

Product Description

The Tri-Flo Subglottic Suction System

Reason for Recall

Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body.

Distribution Pattern

Domestic Distribution: AZ, CA, CT, FL, GA, IA, IL, IN, MA, MD, MN, NJ, OH, OK, SD, TX, WA, WI.

Code Information

Part number: CM28010; Lot/Serial Numbers: 0001158835 , 0001158836, 0001158837, 0001185564, 0001194114 , 0004008255, 0004008256.