Z-0020-2015 Class II Terminated
FDA device recall Z-0020-2015 was initiated by Ion Beam Applications S.A. on Mar 27, 2014 and is designated Class II. Reason for recall: While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the… The recall status is terminated (terminated Dec 10, 2014). Affected quantity: 8.
Recall details
- Product Type
- Devices
- Report Date
- Nov 5, 2014
- Initiation Date
- Mar 27, 2014
- Termination Date
- Dec 10, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8
Product description
Proteus 235 Proton Therapy System for cancer treatment.
Reason for recall
While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.
Distribution pattern
Worldwide Distribution-USA (nationwide) including the states of PA, NJ, IL, WA, and TN, and the countries of Germany, Italy, and Czech Republic.
Code information
PAT.107, PAT.109, PAT.112, PAT.113, PAT.114, PAT.115, PAT.116, PAT.119.