Z-0066-2019 Class II Terminated
Recalled by Johnson & Johnson Surgical Vision Inc
— Milpitas, CA
FDA device recall Z-0066-2019 was initiated by Johnson & Johnson Surgical Vision Inc on May 9, 2018 and is designated Class II. Reason for recall: Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment… The recall status is terminated (terminated Jan 13, 2021). Affected quantity: 9.
Recall details
- Product Type
- Devices
- Report Date
- Oct 10, 2018
- Initiation Date
- May 9, 2018
- Termination Date
- Jan 13, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9
Product description
OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Corporation 510 Cottonwood Drive Milpitas, CA 95035 USA The Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Reason for recall
Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.
Distribution pattern
US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.
Code information
Model: Catalys Precision Laser System, CE 0044: Catalys-U (US version) Serial Numbers: 44065613 - software cOS 5.00.33 install date 4/27/2018 46131618 - software cOS 5.00.33 install date 4/23/2018 46131418 - software cOS 5.00.33 install date 4/30/2018 44107315 - software cOS 5.00.33 install date 5/1/2018 46120916 - software cOS 5.00.33 install date 4/23/2018 46123116 - software cOS 5.00.33 install date 5/2/2018 44106215 - software cOS 5.00.33 install date 5/3/2018 Catalys-I (International version) Serial Number: 44109015 - software cOS 5.00.33 install date 5/1/2018 46124617 - software cOS 5.00.33 install date 5/3/2018
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0066-2019
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Johnson & Johnson Surgical Vision Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Johnson & Johnson Surgical Vision Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Johnson & Johnson Surgical Vision Inc →
Use it in ChatGPT
Get a free API key
How the connector works