Z-0210-2016 Class III Terminated
FDA device recall Z-0210-2016 was initiated by Steris Corporation on Sep 3, 2015 and is designated Class III. Reason for recall: In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close… The recall status is terminated (terminated Jan 22, 2016). Affected quantity: 32 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 11, 2015
- Initiation Date
- Sep 3, 2015
- Termination Date
- Jan 22, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32 units
Product description
AMSCO 400 and AMSCO C Small Steam Sterilizers
Reason for recall
In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close the sterilizer door.
Distribution pattern
US nationwide distribution.
Code information
Serial #'s: 033331205, 033341248, 033351210, 033401221, 033411215, 033451217, 030741306, 031491314, 032381302, 032341304, 033281222, 030621412, 030871412, 030871410, 031751401, 031771425, 031781412, 031781414, 031911436, R020141502, R020151501, R020191502, R020201501, R020261502, R020261504, R020271501, 030341514, R020471501, 030501540, 030651519, R021001503, R021051501