Z-0241-2023 Class II Terminated

Recalled by Steris Corporation — Mentor, OH

Recall Details

Product Type
Devices
Report Date
November 23, 2022
Initiation Date
October 17, 2022
Termination Date
May 28, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 units

Product Description

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001

Reason for Recall

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result

Distribution Pattern

US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia.

Code Information

UDI: 00724995180256 Serial Numbers: 0032-0822-003, 0032-0622-009, 0032-0922-001, 0032-0822-001, and 0032-0082-002