Z-0526-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 24, 2014
- Initiation Date
- April 30, 2014
- Termination Date
- March 16, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32 units
Product Description
Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.
Reason for Recall
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
Distribution Pattern
Worldwide Distribution.
Code Information
Part Number P1003290001